Aortic valve area calculation

Aortic valve area calculation is an indirect method of determining the area of the aortic valve. The calculated aortic valve orifice area is currently one of the measures for evaluating the severity of aortic stenosis. A valve area of less than 0.8 cm² is considered to be severe aortic stenosis.[1][2]

There are many ways to calculate the valve area of aortic stenosis. The most commonly used methods involve measurements taken during echocardiography. For interpretation of these values, the area is generally divided by the body surface area, to arrive at the patient's optimal aortic valve orifice area.

Contents

  • Planimetry 1
  • The continuity equation 2
  • The Gorlin equation 3
  • The Hakki equation 4
  • Real-time three-dimensional echocardiography 5
  • See also 6
  • References 7

Planimetry

Planimetry is the tracing out of the opening of the aortic valve in a still image obtained during echocardiographic acquisition during ventricular systole, when the valve is supposed to be open. While this method directly measures the valve area, the image may be difficult to obtain due to artifacts during echocardiography, and the measurements are dependent on the technician who has to manually trace the perimeter of the open aortic valve. Because of these reasons, planimetry of aortic valve is not routinely performed.

The continuity equation

The continuity equation states that the flow in one area must equal the flow in a second area if there are no shunts between the two areas. In practical terms, the flow from the left ventricular outflow tract (LVOT) is compared to the flow at the level of the aortic valve. In echocardiography the aortic valve area is calculated using the velocity time integral (VTI) which is the most accurate method and preferred. The flow through the LVOT, or LV Stroke Volume (cm3 or cc), can be calculated by measuring the LVOT diameter (cm), squaring that value, multiplying the value by 0.78540 giving cross sectional area of the LVOT (cm2)and multiplying that value by the LVOT VTI(cm), measured on the spectral Doppler display using pulsed-wave Doppler. From these, it is easy to calculate the area (cm2) of the aortic valve by simply dividing the LV Stroke Volume (cm3) by the AV VTI(cm) measured on the spectral Doppler display using continuous-wave Doppler.

SV = 0.785 x Diameter2 x VTI(LVOT)

CSA(LVOT)cm2 = 0.785 x LVOT Diameter

SV = Stroke Volume CSA = Cross sectional area

\text{Aortic Valve Area (cm}^2\text{)} = {\text{LVOT diameter}^2 \cdot 0.78540 \cdot \text{LVOT VTI}\over \text{Aortic Valve VTI}}

Example: An individual undergoes transthoracic echocardiography for the evaluation of a systolic ejection murmur with delayed carotid upstroke noted on physical examination. During echocardiography, the following measurements were made: LVOT diameter of 2 cm, LVOT VTI of 24 cm, and an Aortic Valve VTI of 50 cm. What is the aortic valve area?
Answer: An LVOT diameter of 2 cm gives a LVOT cross-sectional area of, 2 * 2 * 0.78540 = 3.14 cm2. To calculate stroke volume, multiply the cross-sectional area of 3.14 cm2 by the LVOT VTI 24 cm. This gives an LV Stroke Volume of 3.14 * 24 = 75.40 cc. Divide the LV Stroke Volume, 75.40 cc by the Aortic Valve VTI, 50 cm and this gives an aortic valve area of 75.40 / 50 = 1.51 cm2.

The weakest aspect of this calculation is the variability in measurement of LVOT area, because it involves squaring the LVOT dimension. Therefore, it is crucial for the sonographer to take great care in measuring the LVOT diameter.

The measurement using echocardiogram may be inaccurate in cases of Aortic subvalvular stenosis, because there is not a uniform diameter, as assumed during echocardiogram. The viewpoint of the echocardiogram may distort the true diameter of the LVOT in some patients. For verification purposes of the obtained valve area using echocardiogram and doppler measures, especially if the obtained valve area is in the range requiring surgery and cardiac output is low, the Gold standard of left heart catheterization for true hemodynamics should be obtained for validation using the Gorlin formula, so patient does not go for unneeded surgery.

The Gorlin equation

The Gorlin equation states that the aortic valve area is equal to the flow through the aortic valve during ventricular systole divided by the systolic pressure gradient across the valve times a constant. The flow across the aortic valve is calculated by taking the cardiac output (measured in liters/minute) and dividing it by the heart rate (to give output per cardiac cycle) and then dividing it by the systolic ejection period measured in seconds per beat (to give flow per ventricular contraction).

\text{Valve Area (cm}^2\text{)} = \frac{\text{Cardiac Output }(\frac{\text{ml}}{\text{min}})}{\text{Heart rate }(\frac{\text{beats}}{\text{min}})\cdot \text{Systolic ejection period (s)}\cdot 44.3 \cdot \sqrt{\text{mean Gradient (mmHg)}}}

The Gorlin equation is related to flow across the valve. Because of this, the valve area may be erroneously calculated as stenotic if the flow across the valve is low (i.e. if the cardiac output is low). The measurement of the true gradient is accomplished by temporarily increasing the cardiac output by the infusion of positive inotropic agents, such as dobutamine.

Example: An individual undergoes left and right heart cardiac catheterization as part of the evaluation of aortic stenosis. The following hemodynamic parameters were measured. With a heart rate of 80 beats/minute and a systolic ejection period of 0.33 seconds, the cardiac output was 5 liters/minute. During simultaneous measurement of pressures in the left ventricle and aorta (with the use of one catheter in the left ventricle and a second in the ascending aorta), the mean systolic pressure gradient was measured at 50 mmHg. What is the valve area as measured by the Gorlin equation?
Answer: \text{Aortic Valve Area}= \frac{5000 \frac{\text{ml}}{\text{min}}}{80 \frac{\text{beats}}{\text{min}} \cdot 0.33 \text{s} \cdot 44.3 \cdot \sqrt{50 \text{mmHg}}} \approx 0.6 \text{cm}^2

The Hakki equation

The Hakki equation[3] is a simplification of the Gorlin equation, relying on the observation that in most cases, the numerical value of \text{heart rate (bpm)} \cdot \text{systolic ejection period (s)} \cdot 44.3 \approx 1000. The resulting simplified formula is:

\text{Aortic Valve area (cm}^2\text{)} \approx\frac{\text{Cardiac Output} (\frac{\text{litre}}{\text{min}})}{\sqrt{\text{Peak to Peak Gradient (mmHg)}}}

Example: An individual undergoes left and right cardiac catheterization for the evaluation of aortic stenosis. Measurements includes an aortic pressure of 120/60, LV pressure of 170/15, cardiac output of 3.5 liters/minute. What is the aortic valve area?
Answer: The peak gradient between the LV and aorta is 50 mmHg. This gives

\text{Aortic valve area} \approx \frac{3.5}{\sqrt {50}}\approx 0.5\ \text{cm}^2

Real-time three-dimensional echocardiography

See http://circ.ahajournals.org/content/112/7/992

See also

https://www.google.com/search?q=Real-time+three-dimensional+echocardiography&sourceid=ie7&rls=com.microsoft:en-us:IE-SearchBox&ie=&oe=

References

  1. ^ Charlson E, Legedza A, Hamel M (2006). "Decision-making and outcomes in severe symptomatic aortic stenosis.". J Heart Valve Dis 15 (3): 312–21.  
  2. ^ Varadarajan, P; Kapoor, N; Bansal, RC; Pai, RG (2006). "Survival in elderly patients with severe aortic stenosis is dramatically improved by aortic valve replacement: results from a cohort of 277 patients aged >/=80 years.". Eur J Cardiothorac Surg 30 (5): 722–7.  
  3. ^ Hakki A, Iskandrian A, Bemis C, Kimbiris D, Mintz G, Segal B, Brice C (1981). "A simplified valve formula for the calculation of stenotic cardiac valve areas.". Circulation 63 (5): 1050–5.  
This article was sourced from Creative Commons Attribution-ShareAlike License; additional terms may apply. World Heritage Encyclopedia content is assembled from numerous content providers, Open Access Publishing, and in compliance with The Fair Access to Science and Technology Research Act (FASTR), Wikimedia Foundation, Inc., Public Library of Science, The Encyclopedia of Life, Open Book Publishers (OBP), PubMed, U.S. National Library of Medicine, National Center for Biotechnology Information, U.S. National Library of Medicine, National Institutes of Health (NIH), U.S. Department of Health & Human Services, and USA.gov, which sources content from all federal, state, local, tribal, and territorial government publication portals (.gov, .mil, .edu). Funding for USA.gov and content contributors is made possible from the U.S. Congress, E-Government Act of 2002.
 
Crowd sourced content that is contributed to World Heritage Encyclopedia is peer reviewed and edited by our editorial staff to ensure quality scholarly research articles.
 
By using this site, you agree to the Terms of Use and Privacy Policy. World Heritage Encyclopedia™ is a registered trademark of the World Public Library Association, a non-profit organization.
 


Copyright © World Library Foundation. All rights reserved. eBooks from Project Gutenberg are sponsored by the World Library Foundation,
a 501c(4) Member's Support Non-Profit Organization, and is NOT affiliated with any governmental agency or department.